AMI Vapor Series AMI Vapor Series

Category:

Description

Parts

Vapor 200C

Vapor 100C

Vapor 100B

Evaporators

1

1

1

Vapor Sorption Station

1 (1/0.1 Torr)

1 (1/0.1 Torr)

1 (10 Torr)

Micropore Analysis Station

yes

Yes

No

Pressure Sensors

7

5

4

Cold Trap

1

2

2

Preparation Bit

No

( Can be prepared in situ)

1

1

Thermostatic Water Bath

1 (0 ~ 50 oC)

1 (0 ~ 50 oC)

1 (0 ~ 50 oC)

Dewar flask (1L)

2

No

No

Dewar flask (3L)

1

2

2

Mechanical Pump

One corrosion – resisting pump

One corrosion – resisting pump, and One mechanical pump

One corrosion – resisting pump (or One mechanical pump)

  • Preparation of adsorbate by evaporator.
  • Use stainless steel and thermostatic control module as manifold. Constant temperature range: room temperature to 50 o
  • Effective cold trap system prevents vapor from polluting vacuum pump and environment.
  • FFKM (Perfluoroelastomer) O rings for banner sealing performance.
  • Measurements of BET surface area and pore size distribution.
  • Measurements of various gas and steam adsorption isotherms.
  • Testing of isosteric heat of adsorption.
  • Determination of selective adsorption coefficient.

Type

Vapor 200C

Vapor 100C

Vapor 100B

Testing Principle

The static volumetric method

Data Features

BET surface area: ≥ 0.01 m2/g

Mesopore: 2 ~ 500 nm (BJH)

Micropore: 0.35 ~ 2 nm (t-plot, DR, MP,DA, HK ,SF)

Physical Adsorbate

N2, Ar, Kr, CO2

Vapor Adsorbate

Carbon monoxide, ammonia, and a substance that is liquid at room temperature.

(Except corrosive, hyper-toxic, hazardous gas)

Same as Vapor 200C and Vapor 100C, but p0 is greater than 1KPa

 Physical Adsorption Temperature

77K, 87K

Vapor Adsorption Temperature

0~50 oC

Steam Recovery

Cold trap in analysis station

Double cold trap in analysis and pretreatment station

Essential Parameters

Evaporator

Phase of adsobate transforming from liquid to gas by evaporator in room temperature, implements test requirements

Adsorption Pressure

P/P0 is less than 1 (p0 is saturated vapor pressure of adsobate in adsorption temperature)

Vacuum Pump

Special corrosion – resistant vacuum pump, molecular pump

Pressure Sensor

1000 Torr, 10 Torr, 1Torr

Sorption Temperature

-196 oC is provided by liquid nitrogen in low temperature Dewar flask;

0 ~ 50 oC is provided by thermostatic water bath.

Number of Gas Inlet

Providing separate gas intakes according to customer requirements. (Number of intakes is multiples of 4).

Gas Recovery

Effective cold trap system prevents from polluting vacuum pump and environment by vapor.

Reviews