Description
Parts |
Vapor 200C |
Vapor 100C |
Vapor 100B |
Evaporators |
1 |
1 |
1 |
Vapor Sorption Station |
1 (1/0.1 Torr) |
1 (1/0.1 Torr) |
1 (10 Torr) |
Micropore Analysis Station |
yes |
Yes |
No |
Pressure Sensors |
7 |
5 |
4 |
Cold Trap |
1 |
2 |
2 |
Preparation Bit |
No ( Can be prepared in situ) |
1 |
1 |
Thermostatic Water Bath |
1 (0 ~ 50 oC) |
1 (0 ~ 50 oC) |
1 (0 ~ 50 oC) |
Dewar flask (1L) |
2 |
No |
No |
Dewar flask (3L) |
1 |
2 |
2 |
Mechanical Pump |
One corrosion – resisting pump |
One corrosion – resisting pump, and One mechanical pump |
One corrosion – resisting pump (or One mechanical pump) |
- Preparation of adsorbate by evaporator.
- Use stainless steel and thermostatic control module as manifold. Constant temperature range: room temperature to 50 o
- Effective cold trap system prevents vapor from polluting vacuum pump and environment.
- FFKM (Perfluoroelastomer) O rings for banner sealing performance.
- Measurements of BET surface area and pore size distribution.
- Measurements of various gas and steam adsorption isotherms.
- Testing of isosteric heat of adsorption.
- Determination of selective adsorption coefficient.
Type |
Vapor 200C |
Vapor 100C |
Vapor 100B |
Testing Principle |
The static volumetric method |
||
Data Features |
BET surface area: ≥ 0.01 m2/g Mesopore: 2 ~ 500 nm (BJH) Micropore: 0.35 ~ 2 nm (t-plot, DR, MP,DA, HK ,SF) |
||
Physical Adsorbate |
N2, Ar, Kr, CO2 |
||
Vapor Adsorbate |
Carbon monoxide, ammonia, and a substance that is liquid at room temperature. (Except corrosive, hyper-toxic, hazardous gas) |
Same as Vapor 200C and Vapor 100C, but p0 is greater than 1KPa |
|
Physical Adsorption Temperature |
77K, 87K |
||
Vapor Adsorption Temperature |
0~50 oC |
||
Steam Recovery |
Cold trap in analysis station |
Double cold trap in analysis and pretreatment station |
Essential Parameters
Evaporator |
Phase of adsobate transforming from liquid to gas by evaporator in room temperature, implements test requirements |
Adsorption Pressure |
P/P0 is less than 1 (p0 is saturated vapor pressure of adsobate in adsorption temperature) |
Vacuum Pump |
Special corrosion – resistant vacuum pump, molecular pump |
Pressure Sensor |
1000 Torr, 10 Torr, 1Torr |
Sorption Temperature |
-196 oC is provided by liquid nitrogen in low temperature Dewar flask; 0 ~ 50 oC is provided by thermostatic water bath. |
Number of Gas Inlet |
Providing separate gas intakes according to customer requirements. (Number of intakes is multiples of 4). |
Gas Recovery |
Effective cold trap system prevents from polluting vacuum pump and environment by vapor. |
Reviews