Description
Multistandard Capability
ISO 4696-1 Disintegration |
ISO 7215 Relative Reducibility |
ISO 4696-2 Reduction Degradation |
ISO 4698 Free swelling |
ISO 8371 Decrepitation |
IS 10823 Sect.1 B – Static Reduction Degradation |
IS 11292 Relative Reducibility |
|
---|---|---|---|---|---|---|---|
Test Temp. (°C) |
500 | 900 | 550 | 900 | 700 | 500 | 900 |
Reduction Time (min) |
60 | 180 | 30 | 60 | 30 | 60 | 180 |
Gas composition Mix |
A | B | B | B | — | B | B |
Gas flow rate (nl/min) |
20 | 15 | 15 | 15 | — | 20 | 15 |
Sample mass (g) |
500 | 500 | 500 | 18 pellets | 500 | 500 ± 1pc | 500 |
Continuous Weighing system |
Option | Yes | — | — | — | Option | Yes |
Tumbler TB 3000 | Yes | — | Yes | — | — | Yes | — |
Note: SM 4000 can be also equipped with proper hardware and software to run CRI-CSR Test as per ISO 18894, ASTM D 5341 and IS 4023.
Gas Mix Composition % | |||||
MIX | N2 | CO | CO2 | H2 | CH4 |
A | 58 | 20 | 20 | 2 | — |
B | 70 | 30 | — | — | — |
Control Panel
Control panel includes complete top quality control and monitoring instrumentation. Temperature of sample is driven by digital programmer and by multiloop digital controller to obtain best isothermal test condition. A temperature alarm controller assures oven safety cutting off power in case of over temperature.
Automatic inlet of Nitrogen reaction gas are provided according to temperature program. Gas flow control is obtained by digital mass flow meter for reaction gas and by a rotameter for N2 (as option a mass flow meter also for N2 is included).
Manual gas control is also possible for checking purpose. Optional weighing system allows continuous monitoring of sample weight. A Touch screen recorder and data logger, and Ethernet connection, provide local and remote recording of test data.
Specifications
OVEN: | |
Heating elements: | Silicon carbide elements (20 elements) |
Heating Zones: | 5 heating zones |
Max operating temperature: | 1200 °C |
Maximum operating Power: | 20 kW |
External thermocouple: | “K” type (inconel shielded) |
Inner chamber diameter: | 140 mm |
REACTION TUBE: | |
Reaction tube type : | “Double wall” Type B |
Material: | AISI 310 |
Inner diameter: | 75 mm |
Internal thermocouple: | Triple “K” type (inconel shielded) |
GAS CONTROL UNIT: | |
Flow rate measuring devices | Digital Mass Flow Meters |
Reaction Gas flow rate: | Up to 50 nl/min |
Purge Gas composition: | 100% N2 |
Purge gas Flow rate: | Up to 50 nl/min |
SAMPLE WEIGHING UNIT: | |
Max weight: | 30 kg |
Resolution: | 0,1 g |
MAIN POWER: | 380 V 3 Ph + Neutral (20 kW max) |
Vertical electric oven
The electric oven consists of a steel structure placed on a frame and including the silica insulation material. Twenty silicon carbide high temperature heating rod elements are placed horizontally in a stack and are grouped in 5 heating zones to provide separate power control.
Replacement of heating elements is fast and easy with no need of refractory maintenance so that oven life is very long. Inner high temperature chamber is protected by a quartz tube. The oven can be equipped, as an option, with a weighing system to provide continuous monitoring of sample weight during reduction test.
Tumbler system TB 3000
According to ISO 4696-1, ISO 4696-2, IS 10823 and IS 8167, the sample is tumbled after the reaction in the oven. Number of revolutions is pre-set on a digital counter showing remaining revolutions to the end of tumbling. Automatic stop at end of test is provided. The units are protected by a safety cage with switch on the door. Please refer to the Tumbler System product page for further information.
Auxiliary Equipment
REACTION TUBE LIFTING ELECTRIC TACKLE
• For insertion and removing of the reaction tubes
• Vertical electric movement
• Two controlled speeds – low / fast movement
• Manual horizontal movement on rail
• Max. weight 60 kg
• Remote control
• Rail included
• Power 230 V 50 Hz
REACTION TUBE TRIPOD
Model:
• TR 2000 – For CRI-RDI reaction tubes
REACTION TUBE COOLING DEVICES
Reaction tube cooling devices |
• Digital Temperature controller
• Ventilation Fan
• Auto stop at preset temperature
• Power 230 V 50 Hz
Model:
-
CS 2000 – For CRI-RDI reaction tubes
Sample preparation devices
Auxiliary equipment for Sample Preparation can be provided together with main instrumentation.
Available devices are:
- Jaw Crusher for coke
- Sample Splitter
- Testing screen machine (with screening trays)
- Sieving Machine (with sieves)
- Sieves (for manual sieving)
- Thermostatic Oven
- Digital laboratory scale
RB Automazione Catalogue 2017
Download the RB Automazione general Catalogue 2017
Reviews
There are no reviews yet.